Vargas-Sansalvador, Isabel María Pérez de ORCID: 0000-0002-3666-3231, Fay, Cormac ORCID: 0000-0001-9104-5527, Phelan, Thomas, Fernández-Ramos, M.D., Capitán-Vallvey, Luis Fermin, Diamond, Dermot ORCID: 0000-0003-2944-4839 and Benito-Lopez, Fernando ORCID: 0000-0003-0699-5507 (2011) A new LED-LED portable CO2 gas sensor based on an interchangeable membrane system for industrial applications. Analytica Chimica Acta, 699 (2). pp. 216-222. ISSN 0003-2670
Abstract
A new system for CO2 measurement (0-100%) by based on a paired emitter-detector diode arrangement as a colorimetric detection system is described. Two different configurations were tested: configuration 1 (an opposite side configuration) where a secondary inner-filter effect accounts for CO2 sensitivity. This configuration involves the absorption of the phosphorescence emitted from a CO2-insensitive luminophore by an acid-base indicator and configuration 2 wherein the membrane containing the luminophore is removed, simplifying the sensing membrane that now only contains the acid-base indicator. In addition, two different instrumental configurations have been studied, using a paired emitter-detector diode system, consisting of two LEDs wherein one is used as the light source (emitter) and the other is used in reverse bias mode as the light detector. The first configuration uses a green LED as emitter and a red LED as detector, whereas in the second case two identical red LEDs are used as emitter and detector. The system was characterised in terms of sensitivity, dynamic response, reproducibility, stability and temperature influence. We found that configuration 2 presented a better CO2 response in terms of sensitivity.
Metadata
Item Type: | Article (Published) |
---|---|
Refereed: | Yes |
Uncontrolled Keywords: | Carbon dioxide sensor; Gas sensor; Optical sensor; Paired emitter detector-diode sensor; Portable instrumentation |
Subjects: | Engineering > Mechanical engineering Engineering > Electronics Engineering > Electronic engineering Physical Sciences > Chemistry |
DCU Faculties and Centres: | DCU Faculties and Schools > Faculty of Science and Health > School of Chemical Sciences Research Initiatives and Centres > CLARITY: The Centre for Sensor Web Technologies Research Initiatives and Centres > National Centre for Sensor Research (NCSR) |
Publisher: | Elsevier |
Official URL: | http://dx.doi.org/10.1016/j.aca.2011.05.019 |
Copyright Information: | © 2011 Elsevier |
Use License: | This item is licensed under a Creative Commons Attribution-NonCommercial-Share Alike 3.0 License. View License |
Funders: | Ministerio de Ciencia e Innovacion (CTQ2009-11428-C02-01), Direccion General de Investigacion y Gestion del Plan Nacional de I+D+i (CTQ2009-14429-C02-02), Jaunta de Andalucia (Proyecto de Excelencia P08-FQM-3535), Science Foundation Ireland |
ID Code: | 16494 |
Deposited On: | 25 Aug 2011 13:01 by Cormac Fay . Last Modified 28 May 2019 15:40 |
Documents
Full text available as:
Preview |
PDF (A NEW LED-LED PORTABLE CO2 GAS SENSOR BASED ON AN INTERCHANGEABLE MEMBRANE SYSTEM FOR INDUSTRIAL APPLICATIONS)
- Requires a PDF viewer such as GSview, Xpdf or Adobe Acrobat Reader
340kB |
Downloads
Downloads
Downloads per month over past year
Archive Staff Only: edit this record